Design of Analytical Balance for Calibration of Measuring Cup of Observatory Type Rain Gauge with Automatic Pump

Muhammad Wildan Abdulmajid, Kerista Tarigan, Syahrul Humaidi, Marhaposan Situmorang, Perdinan Sinuhaji, Yahya Darmawan

Abstract


The Observation type rain gauge is a manual type rain gauge that requires a measuring cup to measure rainfall in millimeters (mm) of rain. Precise measurement requires calibration of the measuring cup so that the scale reading on the measuring cup can be trusted. Calibration of measuring cups using the mass method related to volume can use an analytical balance. Technological developments can make manual analytical balance calibration automatic by utilizing microcontrollers and pumps in filling the volume. The design of the analytical balance gets results where the average correction value for the 50 ml set point is 0.001ml, the 100 ml set point is -0.760 ml, the 150 ml set point is -0.024 ml, the 200 ml set point is -0.739 ml, and the 250 ml set point is 0.628 ml. The uncertainty value for all set points is ±0.58 ml. The correction value at each set point is different, this is influenced by the value of the measuring cup meniscus, water temperature, and Load Cell sensor. The data generated by the tool can be downloaded through the Arduino IDE application and produced files in Excel.

Keywords


Measuring Cup, Calibration, Analytical Balance, Uncertainty, Arduino IDE, Excel

Full Text:

PDF

References


Cook, R.R. (2002). Assessment of Uncertainties of Measurement for Calibration and Testing Laboratories. NATA. Australia.

JCGM 100: (2008) – Evaluation of measurement data – Guide to the expression of uncertainty in measurement. Joint Committee for Guides in Metrology.

JCGM 200: (2012) – International Vocabulary of Metrology – Basic and General Concepts and Associated Terms (VIM). Joint Committee for Guides in Metrology

ASTM International, (2012). Standard Practice for Calibration of Laboratory Volumetric Apparatus E 542-01. West Conshohocken.

Hogan, Richard. (2015). The Essential Reference Guide to Sources of Uncertainty in Measurement for Every Uncertainty Budget. Isobudgets, Yorktown.

ISO/IEC 17025. (2017). General requirements for the competence of testing and calibration laboratories, 3rd edn. ISO, International Organization for Standarization, Geneva, Switzerland.

Syam, Wahyudin P. (2017). Pengukuran Geometri dan Analisis Ketidakpastian. Nottingham.

Harris, Georgia L., & Miller, Val R. (2017). Selected Procedures for Volumetric Calibrations. NIST.

Wahyudi. (2017). Analisis Aplikasi Sensor Load Cell Sebagai Pendeteksi Berat Buah Pada Sistem Penyorir Buah Otomatis. Skripsi. Palembang

WMO. (2018). Guide to Instruments and Methods of Observation Volume I – Measurement of Meteorological Variables. World Meteorological Organization (WMO-No. 08), Geneva.

M, Diina Qiyaman, Sirenden, Bernadus H., Maftukhah, Tatik, & Eka, Nur Tjahyo. (2018). Karakterisasi Volume Kalibrator Alat Penakar Curah Hujan Menggunakan Metode Gravimetri. LIPI.

Premana, Dian, R.T., Masbah, & Multi, Abdul. (2019). Development of Portable Rain Calibrator Prototype using Submersible Pump. IRSTC.

KAN. (2019). Pedoman Mengenai Kalibrasi Peralatan Volumetrik, KAN Pd-02.08. Komite Akreditasi Nasional.

Kementerian Perdagangan. (2020). Syarat Teknsi Standar Ukuran Metrologi Legal Besaran Volume, No. 124. Direktur Jenderal Perlindungan Konsumen dan Tertib Niaga. Jakarta.




DOI: https://doi.org/10.33394/j-ps.v11i2.7825

Refbacks

  • There are currently no refbacks.




Copyright (c) 2023 Muhammad Wildan Abdulmajid, Kerista Tarigan, Syahrul Humaidi, Marhaposan Situmorang, Perdinan Sinuhaji, Yahya Darmawan

Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 International License.

Creative Commons License
J-PS (Prisma Sains: Jurnal Pengkajian Ilmu dan Pembelajaran Matematika dan IPA IKIP Mataram) p-ISSN (print) 2338-4530, e-ISSN (online) 2540-7899 is licensed under a Creative Commons Attribution 4.0 International License.

View My Stats